Skip to main content

Clino® One Way Premium

Wipe for special requirements

Back to our wipes for special requirements

The Clino® One Way Premium disposable wipe, a microfibre wipe made of 100% polyester, is characterised by very good and gentle cleaning performance, even with oily and dusty contamination. Even though it is made of purely synthetic fibres, the wipe has a sufficient absorption capacity. The chemical resistance is given with acids (peracetic acid), alkalis (caustic soda, nitric acid), bleaching agents, solvents (isopropanol, acetone, benzene compounds). The wipes are primarily used in areas that are controlled purely by particles, but can also be used at any time in microbiological controlled zones. Cutting is done by laser, which simultaneously seals the edges. The wipes are then professionally decontaminated in a cleanroom laundry and packed under ISO 5 conditions. Due to the cleanroom-compatible post-cleaning, the wipes have relatively low levels of ionic and metallic contamination.

Facts

Particle emission (dry)
Particle emission (wet)
Abrasion resistance
Wet cleaning
Dry cleaning
Chemical permanence
Electrostatic behaviour
Softness
Sealed edges
decontaminated
sterile

Special requirements

Clino® One Way Premium

Microfibre wipe, knitted, 100% PES

-

Properties

  • 100% polyester
  • microfibre filament knit fabric
  • laser cut and sealed edges
  • decontaminated in a cleanroom laundry
  • vacuum-packed twice in PE film in a cleanroom
  • also available as gamma-irradiated version on request

 


Advantages

  • chemically resistant
  • very low emission of particles, fibres and extractable substances
  • non-abrasive, no scratching of sensitive surfaces
  • high and residue-free absorbency and absorption of impurities such as dust and oil-based contaminants

Applications

  • effective removal of particulate and organic contamination
  • very suitable for wiping highly sensitive surfaces, e.g. LCD or optical surfaces
  • for polishing and cleaning of product surfaces, e.g. magnetic heads

Product recommen­dation based on cleanroom classes


123456789low dust


ABCDE

Of course, cleanroom wipes used in ISO 5 can also be used in ISO 9, but in this case the economic efficiency and usefulness should be considered.

A 1 to 1 allocation of cleanroom wipes to an air cleanliness class according to ISO 14644-1 is not possible. Recommendations can only be made on the basis of special properties relevant from a cleanroom technical point of view, such as "abrasion resistance" or "particle emission". Users can find further information on this in VDI Guideline 2083 Part 9.2.

Technical data

Properties   Unit of measurementValueTest method
Material     100% PES  
Edge processing     laser cut  
Mass per unit area   g/m2 210 ± 5 ASTM D3776
Standard Test Methods for Mass Per Unit Area (Weight) of Fabric
Wipe weight   g/wipe 19.5 ASTM D3776
Thickness   mm 0.38 ± 0.02 ASTM D1777, ISO 9073-2
Absorptive capacity in water
(Ai) intrinsic
(Ae) extrinsic
  ml/g
ml/m2
n. s.
305
IEST-RP-CC004.3
Section 8.2
Absorbency rate   sec. < 5 IEST-RP-CC004.3
Section 8.2
NVR
Non-volatile residues
IPA based
DI water based
g/m2
g/m2
< 0.8
< 1.5
IEST-RP-CC004.3
Section 7
Particulate residues > 0.5 µm particles/wipe 100 IEST-RP-CC003.4
Fibre residues > 100 fibres/m2< 500 IEST-RP-CC003.4
Organic substances   µg/g n. d. IEST-RP-CC004.3 Section 7
Ionic residues Chloride (ClO2-)
Nitrate (NO3-)
Sodium (Na+)
Sulphate (SO42-)
µg/g
µg/g
µg/g
µg/g
< 4
< 4
< 4
< 4
IEST-RP-CC004.3 Section 7
Organic contaminants silicone oil
amides
D-n-octylphthalate (DOP)
  n. s.
n. s.
n. s.
by FTIR spectrometer
Fourier transform infrared spectrometer
  DimensionsPU per bagPU per carboard boxArt. No.
  12" x 12"
(~ 30 x 30 cm)
150 pieces
900 pieces
(6 x 150)
55430
      n. s. = not specified
n. d. = not detectable
 

Note

The manufacturer's technical data listed above are typical values at the time of publication. They are not specifications. All data are without guarantee, changes are possible without prior notice.
Due to different test methods depending on the respective manufacturers and test laboratories, most of the given values cannot be compared one-to-one with those of other manufacturers/wipes.
Independent tests can be commissioned on request.

To top